Answer :
Final answer:
The dose calibrator constancy test results for 57Co, 99mTc, and 123I fall within the standard of practice limits, considering a 10% margin. However, the result for 201Tl falls just slightly outside the acceptable range.
Explanation:
The standard of practice usually allows for a variation of 10% between measured and calculated values in a dose calibrator constancy test. Let's apply this standard to the numbers you provided:
- 57Co: (56.5 - 57.8) / 57.8 = -0.022, or -2.2%, which is within the acceptable range
- 99mTc: (59.2 - 58.7) / 58.7 = 0.0085, or 0.85%, again within the acceptable measure
- 123I: (46.9 - 51.7) / 51.7 = -0.092, or -9.2%, which, while close to the limit, is still ok
- 201Tl: (47.9 - 53.5) / 53.5 = -0.104, or -10.4%, which falls just outside the acceptable margin
So based on these calculations, the answer to which measured values fall within the limits according to the standard of practice are: 57Co, 99mTc, and 123I, giving you option b.